Abstract
Conventional methods for the determination of major nutrients and trace elements in grass rely on acid digestion followed by analysis using inductively coupled plasma optical emission spectrometry (ICP-OES), which can be both time consuming and costly. Energy dispersive X-ray fluorescence (EDXRF) spectrometry offers a rapid alternative that can determine multiple elements in a single scan. Copper, Mn, Zn, and S in grass samples were determined using EDXRF with a number of different calibration approaches using both empirical standards and the theoretical relationships between concentrations and intensities. Using an existing archive of 467 grass samples of known concentrations, a suite of 30 samples was selected as empirical grass standards to build a calibration set between sample concentrations and EDXRF intensities. The theoretical or standardless approach used the fundamental parameters method to determine element concentrations. To validate the two calibration methods, 59 samples were randomly selected from the same archive and database and analyzed by EDXRF. The measurements of Cu, Mn, Zn, and S were compared with the ICP-OES values using agreement statistics. An excellent correlation was observed between the concentrations determined by EDXRF and ICP-OES (