Abstract
Introduction
Many types of endoscopes are used in clinical diagnosis and treatment. However, the decontamination of these endoscopes is challenging. From 2011 to 2019, the Emergency Care Research Institute has advised that endoscope cleaning is of considerable importance. Reprocessing lapses are regarded as a widespread and ongoing problem, despite the existence of guidelines. 1 , 2 Cystoscopies and ureteroscopies have been linked to outbreaks of infections. 3 , 4 Unfortunately, some healthcare workers do not fully follow the guidelines and protocols necessary to ensure the microbiological safety of endoscopes. There is increasing evidence that infections are transmitted by flexible endoscopes. 5 A bacterial biofilm is defined as a community of microorganisms encased within an extracellular polysaccharide matrix, which adheres to a surface. 6 Bacterial biofilms are associated with a wide range of infections. Notably, more than 60% of endoscopic infections have been caused by bacterial biofilms. 7 A large number of studies have shown that biofilms persist in sterilized endoscopes. 8 Specific cleaning methods are needed remove residual bacteria and reduce the probability of biofilm formation. A recent study of flexible endoscopes (including ureteroscopes) revealed that more than 70% were not sufficiently clean to prevent infection. 9 In the USA, a survey of 67 centers in three states demonstrated that more than 28% of the centers did not strictly follow the recommended guidelines for endoscopic reuse. 10
Quality control circles (QCCs) based on the Deming cycle (also known as the plan-do-check-act [PDCA] model) include 10 steps for ensuring adherence to specific procedures. 11 This study was performed to assess the efficacy of the QCC approach with respect to reducing the formation of biofilms in flexible endoscopes in the urology surgical suite.
Methods
Ethical considerations
All analyses were based on guidelines to improve the cleaning quality of flexible endoscopy, which is a routine procedure. Therefore, the requirement for ethical approval or patient consent was waived by the hospital’s ethics committee.
QCC team
The QCC team comprised two doctors, nine nurses (two head nurses, six urological specialist nurses, and one sensor nurse), one quality control manager, and one engineer. Senior titles, junior titles, and master's degree or above were held by 23%, 21%, and 30% of the team, respectively. Table 1 shows the demographics of the team.
Demographics of the quality control circle team.
Plan
Issue selection
The 13 members of the QCC team voted to select one issue from among five issues generated by circle members, scored according to the following characteristics: urgency, importance, hospital policy, and lap capacity based on the “5-3-1” rule. 12 The most important issue was to reduce biofilm formation in flexible endoscopes by improvement of the qualified cleaning rate, based on the weighted rating method.
Plan and goal establishment
The study was designed based on implementation of the PDCA model in Zhujiang Hospital, Southern Medical University from March 2018 to May 2019. Cleaning processes were examined for three types of flexible endoscopes: flexible video-uretero-renoscopes (Karl Storz 11278VS; Karl Storz, Tuttlingen, Germany), uretero-fiberscopes (Karl Storz 11278A1; Karl Storz), and flexible cystoscopes (HAWK DPG-II; Hangzhou Hawk Optical Electronic Instruments Co., Ltd., Hangzhou, China). Adenosine triphosphate (ATP) detection was used to monitor the efficacy of biofilm removal from flexible endoscopes after the completion of cleaning during the period from 1 March 2018 to 31 June 2018. The internal and external parts of flexible endoscopes were used as sampling sites by means of the flushing and smudge methods, respectively.
In total, 235 operations were examined in the planning period, including 45 involving a flexible video-uretero-renoscope, 93 involving a uretero-fiberscope, and 97 involving a flexible cystoscope. The results showed that the ATP monitoring pass rate was 50%, based on a criterion of concurrent external and internal ATP monitoring results less than 500 relative light units. Based on the results of a questionnaire checklist for cleaning flexible endoscopes before establishment of the QCC, the cumulative frequency of four dominant factors (monitoring, pre-treatment, cleaning, and leak detection; all determined using Pareto’s principle
12
) was 81.18%. These four factors were regarded as the focus of efforts to prevent the formation of biofilms in flexible endoscopes in the urology surgical suite, prior to sterilization. The results of Pareto analysis before improvement are shown in Figure 1a. This study set the following goals for removal of biofilm from flexible endoscopes:
Goal value = value before improvement +(1–value before improvement) × lap capacity × improvement focus × 100% Growth rate = (goal value – value before improvement)/value before improvement × 100%

Pareto analysis before and after improvement. a: before improvement; b: after improvement.
The ATP monitoring pass rate before improvement was 50%, the lap capacity (i.e., according the “5-3-1” rule, grading registration was divided into three levels: 12 5 = solve problems by themselves, 3 = one department must cooperate; and 1 = many departments must cooperate; lap capacity =average score/5 × 100%) was calculated to be 82.23%, and the improvement focus (i.e., cumulative frequency based on the goal of solving the first 80% of major problems 12 ) was 81.18%. Therefore, the team planned to increase the pass rate from 50% to 83.38% by January 2019, which constituted a 38.41% growth rate.
Cause analysis
During “brainstorming” sessions, fishbone diagrams were generated by the team members (Figure 2). For monitoring of four factors including monitoring unqualified, pretreatment unqualified, cleaning unqualified, and side leakage unqualified, 32 terminal causes were proposed (Table 2) from the following five aspects: people, machines, materials, methods, and environment. Team members scored the terminal causes. In accordance with Pareto’s principle, terminal causes with scores >52 were extracted. Finally, seven causes that might affect endoscope reprocessing were identified: lack of consistent cleaning personnel, lack of professional training, lack of a traceability system, lack of pre-treatment wipes, lack of specified leak detection process, lack of detailed cleaning, and lack of biofilm removal monitoring process. To identify the “root causes,” the team generated new follow-up checklists. For example, to affirm the lack of pretreatment wipe application, the team developed an audit standard to assess whether enzyme-containing gauze towels were used for pre-treatment. The audit tool was a custom clinical checklist of pre-treatment tasks to be performed after the use of a flexible endoscope; the audit method comprised case-by-case traceability. Audits revealed that all seven of the above suggested causes constituted root causes. Table 3 depicts the root cause analysis.

Fishbone diagram. a: Why was the qualified rate of monitoring low?; b: Why was the pass rate of leak detection low?; c: Why was the qualification rate of pretreatment low?; d: Why was the qualified rate of cleaning low?
Terminal cause analysis.
Root cause analysis.
Solution formulation
In accordance with the principles of “5W1H” (i.e., Why, What, Where, When, Who, and How), the team developed countermeasures. These countermeasures consisted of four categories: creation of an evidence-based process for removal of biofilm from flexible endoscopes, optimization of endoscopic traceability management, introduction of effective pre-treatment wipes, and development of a standardized training program.
Do
Creation of an evidence-based process for removal of biofilm from flexible endoscopes
Prior to establishment of the QCC, there were three relevant root causes for this removal process: lack of a specified leak detection process, lack of a detailed cleaning process, and lack of a biofilm removal-related monitoring process. From 30 November 2018 to 30 December 2018, the team established a process for removal of biofilm from flexible endoscopes, using the following three steps. First, the team reviewed and summarized guidelines13,14 regarding the use of flexible endoscopes, then conducted an evaluation of guidelines for implementation in our hospital. This evaluation comprised quantitative assessment of feasibility, suitability, significance, and effectiveness by three doctors and two nurses from our hospital. Grading registration was divided into four levels: 4 = strongly agree, 3 = agree, 2 = disagree, and 1 = strongly disagree. In accordance with the usability survey results, all evaluated guidelines were implemented. Second, based on the principles of evidence, professional judgment, and situational analysis, the reviewed guidelines were converted into a process for monitoring the removal of biofilm from flexible endoscopes and an accompanying evaluation index that could be used by the surgical endoscopy center in our hospital. The timing and staff in charge of the monitoring process were discussed. Third, combined with the above guidelines and experience in clinical practice, the monitoring index, monitoring method, and timing of flexible endoscopic biofilm removal were determined.
Optimization of endoscopic traceability management
With respect to the lack of consistent cleaning personnel, the team discussed determination of clear job responsibilities to ensure identification of consistent cleaning personnel. The team defined the contents of traceability management and enabled traceability through modifications of the warehouse management software system.
Introduction of effective pre-treatment wipes
Through literature review and market research, team members identified suitable multi-enzyme pre-treatment wet wipes. All literature reviews were performed using the PubMed, ScienceDirect, and CNKI databases during the period from 1 January 2009 to 31 December 2019. The key words used in literature review were “pretreatment,” “endoscopy,” and “detergent.” Pretreatment was confirmed to improve the effects of cleaning and reduce the formation of biofilm on the surface and within the conduit of flexible endoscopes. 13 The market research involved brief introductions of potential products into the hospital. Because some Gram-negative bacteria undergo cell division every 20 to 30 minutes, all reprocessing steps must be completed rapidly prior to the onset of bacterial growth and contamination on the endoscope surface. 15 The team determined that one-time use wipes should be used immediately after operation under the supervision of the circulating nurse in the operating room.
Development of a standardized training program for removal of biofilm from flexible endoscopes
The research team sent a nurse to the National Endoscopy Conference to receive detailed training. On the basis of theoretical training and interpretation of the guide, one-on-one training was carried out for removal of biofilms from flexible endoscopes. Before and after training, all nurses completed assessments of their knowledge regarding biofilm removal.
Check
The flexible endoscope cleaning pass rate was investigated in the urology surgery suite during the period from 1 January 2019 to 12 May 2019. ATP values were compared before and after establishment of the QCC.
Act
The flowchart was standardized for improvement and renewed at 1-year intervals.
Effect confirmation
The evaluation indices for tangible outcomes included achievement and improvement indexes with the following equations:
Achievement Index = (value after improvement – value before improvement)/(goal value – value before improvement) × 100% Improvement Index = (value after improvement – value before improvement)/value before improvement ×100%
New Pareto analyses were repeated to determine whether factors had changed, with respect to unqualified cleaning of flexible endoscopes, after initial improvement.
For intangible outcomes, evaluations were made regarding problem-solving ability, sense of responsibility, communication and coordination, team cohesiveness, self-confidence, and approach to quality control. The activity growth value was used to assess intangible outcomes, such that positive values indicated improvement and negative values indicated deterioration. Each aspect was assessed by QCC members through self-evaluation; the highest possible score was 5 points and the lowest possible score was 1 point, as previously described.
16
Scores were recorded using the following equation:
Results
Tangible outcomes
From 1 January 2019 to 12 May 2019, 235 operations were analyzed, including 48 involving a flexible video-uretero-renoscope, 95 involving a uretero-fiberscope, and 92 involving a flexible cystoscope. The ATP monitoring pass rate (i.e., qualified rate) increased from 50% to 85.29% after establishment of the QCC. According to the calculation formula, the achievement index was 105.72%, while the improvement index was 70.58%. Table 4 shows that the cumulative frequency of the following new dominant factors was 80.77%, based on repeated Pareto analysis: rinse, transport, counts, classification, recording. The results of Pareto analysis after improvement are shown in Figure 1b. Continuous quality improvement showed that the ATP monitoring pass rate among 150 additional operations (April 2019 to October 2019) increased from 85.29% to 90%.
Factors associated with biofilm removal: relationships with qualified rates of flexible endoscopes before and after establishment of the quality control circle.
QCC, quality control circle.
Process for removal of biofilm from flexible endoscopes
The flowchart depicting flexible endoscope reprocessing in the urology surgical suite prior to establishment of the QCC is shown in Figure 3a; reprocessing after establishment of the QCC is shown in Figure 3b. Careful manual cleaning was a prerequisite for further cleaning and disinfection in automated endoscope washer-disinfectors. Artificial cleaning, especially the basis for washing procedures for mechanical cleaning, was considered an irreplaceable step for mechanical cleaning. The process of manual cleaning is shown in Figure 4. Leak detection was an important step. This study established three-leak detection, as shown in Figure 5. In addition, the QCC team established diagrams to help nurses understand the standardized process. A diagram of the manual cleaning process for flexible cystoscopy is shown in Figure 6.

Reprocessing flowcharts of urology flexible endoscopy before and after establishment of the QCC. a: Cleaning comprised six steps before establishment of the QCC; b: Cleaning comprised 14 steps after establishment of the QCC.

Flexible endoscope manual cleaning process.

Leak detection implementation process.

Diagram of manual cleaning process for flexible endoscopes in the urology surgical suite.
Intangible outcomes
After establishment of the QCC, considerable improvements were noted in problem-solving ability, sense of responsibility, coordination, communication and self-confidence, team unity, enthusiasm, and quality control practices and harmony. Specifically, the activity growth values were 1.5 for ability to solve problems, 1.3 for sense of responsibility, 1.3 for communication and coordination, 1.5 for confidence, 1.3 for cohesion, 1.5 for initiative, 2.0 for approach of QC, 1.5 for harmonious degree, 1.4 for language skills, 1.2 for sense of honor, and 1.4 for personal qualification.
Discussion
Endoscopic reprocessing is a multi-stage process. Ten steps (i.e., pretreatment, transfer, leak detection, hand cleaning and self-cleaning, rinsing, monitoring, disinfection/sterilization, final rinsing, drying, and storage) are included in the most recent domestic and international endoscopic reprocessing guidelines.13,14 The European Society of Gastrointestinal Endoscopy, European Society of Gastroenterology and Endoscopy Nurses and Associates, and American Society for Gastrointestinal Endoscopy have emphasized that cleaning is the most important step in the reprocessing process, but that disinfection or sterilization alone is insufficient.17,18 Considering that biofilms are irreversible after they have formed, this study aimed to reduce the biofilm formation in flexible endoscopes by improving the quality of cleaning protocols prior to disinfection/sterilization of flexible endoscopes. The results showed that the pass rate of biofilm removal from flexible endoscopes increased from 50% to 85.29%, indicating that establishment of the QCC was effective for improving the flexible endoscope cleaning process.
The evidence-based countermeasures implemented in this study were developed to enhance cleaning of flexible endoscopes in the urology surgical suite. The “Technical regulations for cleaning and veal of the flexible endoscopes” 15 was the main basis for the development of cleaning and monitoring processes in this study; this set of technical regulations adheres to China's existing standards and norms, as well as international requirements for cleaning and detoxification of flexible endoscopes. In present study, we developed standard processes for general use of flexible endoscopes, including manual cleaning, leak detection, and monitoring.
Previous studies have suggested that improving the awareness of cleaning personnel through regular education and competency assessments is an important strategy for prevention of reprocessing errors. 16 In the present study, all endoscopy nurses were required to pass a regular assessment regarding reprocessing of flexible endoscopes. The results showed an increased quality of reprocessing staff, as well as a reduction of >30% regarding the risk of infection. 17 However, a limited number of specialist endoscopy nurses were responsible for cleaning all endoscopes in our hospital. This approach was implemented to improve the quality of the cleaning process, which ensured consistency.
In conclusion, we established a QCC that implemented a new cleaning process and relevant monitoring. Establishment of this QCC led to improvement of the flexible endoscopy cleaning pass rate. The literature suggests that a conventional assessment and multidisciplinary team microbial monitoring method should be carried out for flexible endoscope reprocessing. 18 Notably, only one method of ATP monitoring was used in this study. During the quality improvement process follow-up analysis, the QCC team will enhance their assessments of the methods used for flexible endoscope reprocessing.