Abstract
Purpose
Elastomeric pumps are used to infuse a 46-h fluorouracil protocol and patients are asked to visually inspect the pump daily. The pump has a variability of ±10% and there are additional patient variables that can increase this. The feasibility of weighing the pump rather than a visual inspection along with the secondary objective to confirm the pump’s variability in real world conditions was undertaken.
Methods
Empty pumps were weighed using both pharmacy and kitchen scales. Pumps upon completion of the 46-h infusion were also weighed using both pharmacy and kitchen scales.
Results
The kitchen scale was as accurate as the pharmacy grade scale. Disconnected pumps showed the expected variability from using these infusor pumps along with a few showing greater variability likely due to patient variables.
Conclusion
Weighing pumps appears to be feasible both at the pharmacy and home level. Next steps would be to weigh pumps during the infusion to validate an alternate method to simple visual inspection for patients to confirm proper infusing of the pump at their home.