Right at the beginning of the COVID-19 pandemic, the Copenhagen Emergency Medical Services (CEMS) developed a digital diagnostic device to assess symptoms of infection [1]. In just three days, this device was launched in the Capital Region of Denmark. A week later, the device was implemented nationwide in Denmark and was used more than 90,000 times in its first week and almost 150,000 times in the second week. The purpose of the device was presented as twofold [1–4]: (a) to help individual citizens in assessing whether symptoms they experienced were potentially COVID-19 related and to advise them when and where to seek further medical assistance; and (b) to reduce the number of calls to public health-care helplines. Immediately, politicians and administrators described this digital diagnostic device in press releases as ‘a new digital solution from the Regions’, which ‘has been a great success’ [4,5]. However, frequent use of the device may not equal alleviation of pressure on the helplines; ‘see a doctor’ is indeed likely advice provided by the device. If high usage is not an adequate criterion of success, what would be? And should such a device by default be considered low risk?
With a rapid turn to digital solutions in a pandemic age, there is a need to move beyond the hype [6–8] and base application of possibilities on clear criteria for evidence of effect. This paper reflects on the uncertain aspects of the performance, safety risks and accountability of the digital diagnostic device in order to draw such lessons for the future.
Digital devices in health care and the absence of evidence
CEMS initiated the development of the device on 12 March 2020, and policy support and research grants have allowed for an ongoing upgrade of the device [9]. Therefore, its concrete content elements have changed over time. With support of their software provider – Microsoft – CEMS designed the device as a simple decision tree (Figure 1) inspired by the decision trees normally used in their telephone assessments of symptoms [1]. The helplines are the entry point into the primary and secondary health-care services in Denmark outside of regular GP office hours.

The decision tree of the Danish COVID-19 ‘chatbot’, by the Copenhagen Emergency Medical Services (CEMS). The figure is based on the first version of the launched device and the decision tree published by CEMS elsewhere [1].
The initiative represents a digital response to the challenges of diagnostics and triage. When patients facing serious illness decide which action to take, it can have the gravest implications. They must make the right choice. However, for digital devices, clinical standards as known from evidence-based medicine (EBM) are not in place [10]. Unlike drugs, software must continuously be developed and prove its efficiency through use. In response to our inquiries, we have been informed that this particular device was not subject to EU regulations of medical devices, as it was declared not to be ‘intended by its manufacturer to be used specifically for diagnostic or therapeutic purposes’ [11]. The Food and Drug Administration (FDA) responded that they would categorise the device as ‘lower risk’ for which the FDA does not intend to enforce requirements under the FDA&C Act at the moment [12].
The device was also presented as a ‘chatbot’, although it does not conduct an actual written or oral conversation with the user, which usually defines chatbots [13,14]. The device is more accurately classified as a computerised diagnostic decision support, widely known as a symptom checker [15]. These are typically available online or as apps [16]. Symptom checkers have been used in other countries to manage COVID-19 [17]. So far, however, uncertainty prevails about their effects [16,18]. A systematic review found variation between symptoms checkers but relatively strong evidence that they are inferior in triage compared to health professionals [18]. They are mostly more cautious, which should work against the goal of minimising pressure on helplines. The review found little evidence of cost-effectiveness and patient compliance [18].
Lessons for future digital devices
With little evidence and high levels of urgency, it is important to help developers to be better prepared for digital responses to pandemic threats. Since diagnostic uncertainty is at the heart of triage function, we draw on the taxonomy of uncertainty in health care by Han et al. [19] and combine it with the principles of EBM and the EU and US regulations of medical devices.
Based on the available information also graciously offered by the developers about the purposes, design and success of the device, the following elements contain lessons for the future:
The stated criterion of success is ‘frequent use’, but this is insufficient when a device intervenes in the diagnostic process. Criteria should be based on assessments of risks and benefits that includes long-term consequences – also for other affected actors. In this case, the appropriateness of recommendations should be a criterion of success.
The scope of the device is limited to detection of COVID-19 (Figure 1). Nonetheless, users may be experiencing symptoms of other potentially acute, fatal and relatively easily treatable diseases, such as meningitis or acute coronary syndrome, which are not appropriately assessed in the device. During a pandemic, it is important not to expose users to the risk of overlooking symptoms from severe diseases not related to COVID-19. Thus, interventions such as this device should not by default be considered low risk.
The device collects minimum feedback from users in a form of comments and answers about their demographic. Since software should prove its effect through use, it is important to design it with a plan for appropriate data collection to document achievement of the stated criteria of success. This device did not collect sufficient data to evaluate its effect. Therefore, the impact of the device should not be considered low risk.
Clear allocation of responsibilities could stimulate additional critical thoughts during the development phase and provide users with appropriate contact points in case of concerns or challenges. Accountability may prompt politicians, administrators and health-care workers to ensure the quality control of unregulated medical devices. Finally, accountability has a symbolic value with respect to maintaining public trust in pandemic situations (and clinical matters in general).
Despite these concerns, it is remarkable how quickly the device was developed and implemented, especially at the beginning of the outbreak in Denmark, when every institution and workplace were in a state of emergency. Further, CEMS cannot be accountable for the many uncertain aspects of a new disease, including symptoms, prognosis and treatment. Finally, if CEMS is forced to take action on overloaded helplines, they cannot be required to meet the same demands that commercial devices do or guidelines that takes years to produce and formulate. The device should be seen in this specific context, which was – and is – extraordinary.
However, based on the mentioned lessons, a device meant to substitute triage for citizens should not by default be considered low risk. In a pandemic age dependent on digitalisation, developers, initiators and health authorities could take advantage of the overall principles of EBM and medical device regulations (Table I). The suggested step-wise approach does not require many resources. Sometimes you have to build the boat while sailing [20,21], but it only sustains the need for gathering evidence along the way too, while taking into account how privacy concerns and user agreements may significantly hinder or bias the collection of the aimed evidence in practice. We therefore need to update existing scientific and regulatory principles and facilitate their easy use in order to ensure that future devices have the intended performance and safety.
Suggested approach before implementation of a non-regulated digital diagnostic device.
The information provided by the approach outlined above should be publically accessible along with the device to inform users and other affected actors. The approach is based on principles of evidence-based medicine, the EU and US regulations of medical devices and the taxonomy of uncertainty in health care by Han et al. [10,12,19].