Abstract
Introduction
Cancer has been an increasingly serious threat to human health worldwide. 1 About half of patients are estimated to receive radiotherapy at some point after a diagnosis of cancer. 2 Image-guided radiation therapy (IGRT) has increasingly been recognized as a necessary part of precision radiotherapy. The flat-panel kV imaging device integrated into linear accelerator is the most commonly used in room image guidance method. It is capable of acquiring both 2D radiographs and 3D-CBCT images. 3 Recently 4D-CBCT has been developed to address the significant respiratory motion artifacts in 3D-CBCT when imaging the thoracic region. Using 4D-CBCT, multiple phase-based CBCT datasets are generated which characterize the tumor motion at various phases of the breathing cycle. The respiration-correlated breathing signal for 4D-CBCT reconstruction is extracted from the CBCT projection data itself, eliminating the need for an external surrogate or implanted fiducial markers. 4 4D-CBCT has been progressively integrated into the lung cancer SBRT workflow. 5
The imaging dose of 4D-CBCT is currently an issue concerning researchers and users. 6 There are three reasons for the difference in imaging dose between 4D and 3D CBCT, which are explained here using the x-ray volume imaging (XVI) system (Elekta, Stockholm, Sweden) as an example. Firstly, more projections are required to obtain sufficient respiratory motion information, and therefore the acquisition time for 4D-CBCT (on the order of 3-4 min) is significantly longer than for 3D-CBCT (approximately 1 min). 4 Secondly, the precision requirements of lung cancer SBRT treatment have resulted in institutions acquiring 2 to 3 4D-CBCT datasets for each fraction. The pre-treatment 4D-CBCT is used to measure and correct any misalignment of the time-weighted mean tumor position, while the in-treatment (or post-treatment) 4D-CBCT is used to assess any intrafraction patient movement. Some institutions use an additional pre-treatment 4D-CBCT scan to validate the correction applied. 5 Thirdly, during in-treatment 4D-CBCT scan, the XVI flat-panel continues to acquire images at a constant rate of 5.5 frame/s, while the gantry rotation speed depends on the treatment plan and may change constantly. As a result, the projection angles may be uneven and the imaging dose may differ from the pre-treatment 4D-CBCT.
For the above reasons, the imaging dose of 4D-CBCT may be significantly greater than that of 3D-CBCT. Researchers have made efforts to reduce the imaging dose of 4D-CBCT by using fewer projections for reconstruction, 6 but this is not yet clinically available. Therefore, understanding the magnitude of this dose is important in order to minimize its risk. Currently, imaging dose is not taken into account in radiotherapy treatment planning, which may lead to lower dose estimates for OARs.
Monte Carlo (MC) methods were the recommended dose calculation algorithms for kV imaging dose by AAPM TG report 180. 3 Many researchers have evaluated the imaging dose of 3D-CBCT. 7 Some previous studies have built models of the XVI system using MC methods8-10 or treatment planning systems 11 and verified them with measurements in phantoms. Zhang et al evaluated the concomitant imaging dose and dose parameters to organs at risk (OARs) in 3D-CBCT guided thoracic radiotherapy using a Varian on-board imager (OBI), using MC methods. 12
Fewer studies evaluated the imaging dose of 4D-CBCT. Nakamura et al evaluated the imaging dose and dose parameters to OARs from the 4D-CBCT module integrated in the Vero4DRT system (Mitsubishi Heavy Industries Ltd, Tokyo, Japan, and Brainlab AG, Feldkirchen, Germany). 13 Yuasa et al used MC simulation to estimate the organ equivalent dose and effective imaging dose for 4D-CBCT performed with the OBI system, and evaluated the excess absolute risk of secondary cancer incidence. 14 Thengumpallil et al analyzed the 4D-CBCT imaging dose of the XVI system using phantom measurements. 15 To the best of our knowledge, no previous study has used the MC method to evaluate the imaging dose from 4D-CBCT performed with the XVI system, let alone from in-treatment 4D-CBCT scans.
The aim of this study is to evaluate the imaging dose of pre-treatment and in-treatment 4D-CBCT in patients treated with thoracic radiation therapy and its impact on OAR dose parameters based on MC calculations. First, geometric model of a CBCT system was built for MC calculations. The model was calibrated using absolute dose measurement and verified using percentage depth dose (PDD) and profiles measurements in water phantom. Subsequently, imaging doses from pre-treatment and in-treatment 4D-CBCT scans were calculated for 8 lung cancer patients, and the following were evaluated: (1) Dose distribution in patient CTs and dose parameters of typical OARs from pre-treatment and in-treatment 4D-CBCT scans. (2) The relationship between OAR dose parameters from in-treatment 4D-CBCT scans and beam MU or number of frames. (3) The increases in the OAR dose parameters when the imaging dose was added to the planned dose, compared with the OAR dose parameters of the planned dose alone.
Materials and Methods
Ethics Approval and Consent to Participate
The study was approved by the institutional review board of National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital & Shenzhen Hospital in Shenzhen, China, on 28 February 2024 (approval number: JS2024-2-1). Participants provided written informed consent for agreeing the use of personal medical data prior to participating in the study. All patient details were de-identified in this study. The reporting of this study conforms to STROBE guidelines. 16
Patient Data
Simulating 4DCT images and CBCT log files for 14 lung cancer patients previously treated in 2024 at our hospital were randomly selected for this retrospective study according to the approval of the local ethics committee. All patients were used for evaluating the correlation of number of kV frames acquired with in-treatment scan as a function of treatment beam monitor units (MUs). Eight of the patients were randomly selected for evaluating the concomitant imaging dose from 4D-CBCT scans (Table 1). Of the 8 patients, 4DCT images were acquired in 3 patients during free breathing (FB) and average intensity projection (AIP) CTs were used for MC calculations. For the other 5 patients, 3DCT images were acquired during deep inspiration breath hold (DIBH) achieved using the Active Breathing Coordinator (ABC) system (Elekta, Stockholm, Sweden) and were used for MC calculations. All patients were treated on an Infinity linear accelerator (Elekta, Stockholm, Sweden) with 2 or 3 volumetric modulated arc therapy (VMAT) beams with 6 MV flattening filter free (FFF) energy, arc length of 360° and MUs ranging from 505 to 2542 per beam. Since the in-treatment 4D-CBCT images were acquired during delivery of the first treatment beam as stated in our clinical protocol, the first treatment beams from each patient plans were used for evaluation in this study.
Basic Patient Information.
CBCT System and 4D-CBCT Feature
This study evaluated the XVI system version 5.0.7.1-b2 integrated with the Infinity linear accelerator platform. The XVI system consists of a kV x-ray source (70∼150 kVp) with an amorphous silicon/cesium iodide flat-panel detector. It is capable of imaging in either radiographic, fluoroscopic, 3D-CBCT or 4D-CBCT modes. Volumetric images can be acquired with three different field-of-views (FOVs) at the isocenter, namely small (S), medium (M), and large (L), each using different collimators. The two filter options are F0, which uses no filter, and F1, which uses a bow-tie filter. Both collimators and filters are inserted in front of the kV source and can be changed manually.
4D-CBCT scans for lung patients in our hospital are usually performed with S20 collimator and F1 filter. For both pre-treatment and in-treatment 4D-CBCT scans, the acquisition preset parameters were set to 120 kV, 20 mA and 16 ms per frame. These settings were used in the modelling of the CBCT system. A pre-treatment 4D-CBCT scan has an arc length of 360°. It takes approximately 3 min and acquires 1000 frames. For in-treatment 4D-CBCT scans, the imaging panel acquires images at a constant rate of 5.5 frames/s, and the gantry movement is determined by the treatment plan delivery. In the case of this study, all patients were treated with VMAT beams with an arc length of 360°, so the in-treatment 4D-CBCT scans all have an arc length of 360°. An additional AcquisitionInterval parameter was defined and set to 0.1°. During the in-treatment imaging process, when the gantry does not move fast enough and thus the angle between two frames is less than the AcquisitionInterval, a kV source off projection image would be taken instead to avoid redundant images. Table 2 shows details of the imaging protocols for pre-treatment and in-treatment 4D-CBCT.
Comparison of the Imaging Protocols Between Pre-Treatment and in-Treatment 4D-CBCT.
Monte Carlo Code and Simulation Configuration
The TOPAS MC code 17 is an advanced and user-friendly extension to Geant4.18-20 TOPAS version 3.7 with Geant4 version 10.06.p03 was used in this work. The physics module “g4em-standard_opt0” was used. The production threshold for all particles was taken as 10 keV following a previous work. 9 Particle induced x-ray emission (PIXE) was activated in all processes of the simulations.
CBCT System Modeling
The modeling of the CBCT system, including the x-ray tube and the collimation system, followed previous studies.8-10 A phase space file was acquired at the exit plane of the F1 filter with 2 × 1012 incident electrons in the x-ray tube. This phase space file was used for further calculations in phantoms and patient CTs.
Percentage Depth Dose and Lateral Profiles Measurements
PDD and lateral profiles measurements in water were carried out using a computer-controlled scanning phantom Beamscan (PTW) with a waterproof PTW 31021 ionization chamber of 0.07 cm3 nominal active volume. Another PTW 31021 ionization chamber was used as field reference detector in all measurements. PDD and lateral profiles at 1 cm depth were acquired using the WaterTankScans software version 4.2 (PTW) in step mode. The step size varied from 0.2 cm to 1 cm for PDD measurement, and was 0.5 cm for lateral profile measurements.
These measurements were taken using the CBCT system in a stationary gantry position at 270°, with the kV source at 0° for vertical irradiation. The source to surface distance (SSD) was set to 100 cm. The kV beam was delivered 300 frames, with nominal 20 mA and 16 ms per frame.
Absolute Dose Calibration
Absolute dose measurement was carried out with the PTW 30013 Farmer chamber calibrated for absolute dose in kV x-ray according to IPEMB code of practice.
21
The measurement was carried out at the water phantom surface and 30 cm SSD. The dose to water at the water phantom surface was measured using the following equation
The same experimental conditions described above were reproduced with TOPAS MC code. A calibration factor F was then calculated using the following equation:
Validation of the CBCT Model Using Heterogeneous Phantoms
Two different combinations of water equivalent (RW3, PTW, Germany) slabs and tissue mimicking material slabs representing cortical bone (SB3) and lung tissue (LN-300) (SunNuclear, USA) were used as heterogeneous phantoms to validate the accuracy of dose calculations. The lung phantom consisted of 2 cm of RW3 slabs, a 6 cm LN-300 slab and 8 cm of RW3 slabs from top to bottom, and the measurement depths were 1 cm and 9 cm. The bone phantom consisted of 2 cm of RW3 slabs, a 1 cm SB3 slab and 8 cm of RW3 slabs from top to bottom, and the measurement depths were 1 cm and 4 cm. The irradiation setup of the phantoms was the same as for the PDD and profiles measurements. The kV beam was delivered 1000 frames, with nominal 20 mA and 16 ms per frame. Dose measurements were carried out using LD-V1 films (Ashland, USA), and each measurement was repeated for 3 times. The use and calibration of the film generally followed the methodology of a previous study. 22 The phantoms were CT scanned and the MC calculated doses using the same setup were compared with the measurements.
Dose Calculation in Water Phantom and Patient CTs
A water cube with dimensions of (80 cm)3 was used for dose calculations in water. For PDD and lateral profile calculations, the voxel size was set to (0.5 cm)3. When calculating absolute dose in ionization chamber, a geometry model of the PTW 30013 ionization chamber was built in order to calculate the dose collected in the active volume. Detailed descriptions of the geometries were found in previous papers 23 and provided by the manufacturers.
For 4D-CBCT dose calculations in patient CTs, the gantry angle of each frame was extracted from the CBCT log file, transformed into the kV source angle (ie 90° was added), and then written into the MC code using a script. The kV source angles were nearly equidistant for pre-treatment scans and non-equidistant for in-treatment scans, as the latter depended on the treatment plan. The kV source angle of each frame was used as the angle of the phase space source. Using the time feature system of TOPAS, the dose in patient CTs was calculated per frame with a time-dependent kV source angle and then automatically summed to obtain the total 4D-CBCT imaging dose. The isocenter of the kV beam was set to the same value as that used for treatment planning. For in-treatment scans, only frames with the kV source on were used for dose calculations. The CT HU to material converter was implemented following the technique developed by Schneider. 24 The voxel size was set to (0.3 cm)3, the same as the treatment planning dose calculation grid size.
In our institution, 4D-CBCT is most commonly used in SBRT treatments with 5 fractions, and occasionally in treatments with 10 or more fractions. For each treatment fraction, a pre-treatment and an in-treatment 4D-CBCT scan are usually used. To account for the usually ignored impact of imaging dose on OAR parameters, the imaging dose of all the 4D-CBCT scans used throughout the treatment was added to the planning dose. The single-fraction imaging dose was multiplied by the number of fractions to obtain the total imaging dose for the multi-fraction treatment. The OAR parameters of the planning dose with imaging dose were then compared to those of the planning dose alone to determine the increase in OAR dose parameters induced by 4D-CBCT scans. The sum of the planning dose and the imaging dose and the evaluation of the OAR dose parameters, including mean lung and heart doses, maximum spinal cord, esophagus and ribs doses, lung V20 (percentage volume of the whole lung receiving ≥ 20Gy) and V5 (percentage volume of the whole lung receiving ≥ 5Gy), were performed using MIM version 7.3.2 (MIM Software Inc., Cleveland, OH, USA).
To validate the AIP-derived imaging dose calculation, we used a 4D dose calculation method and compared the results with those calculated using AIP CTs. For each kV frame, it is assumed that the x-ray source would irradiate the ten phases of the 4DCT image sets evenly. This is a reasonable assumption, given that the CBCT sampling frequency (5.5 frames per second) is much higher than the breathing frequency (12-20 cycles per minute). Under this assumption, the 4D dose was calculated as the average of the doses calculated on all the individual phases of the 4DCT simulation scan using the kV frames at different angles, without considering the time dependence of the delivery sequence. Deformable registrations were performed between each 4DCT phase and the AIP CT. Then, the dose calculated for each phase could be deformed to the AIP CT, and the average dose could be computed.
Statistical Analysis
All statistical analysis were performed with IBM SPSS V22 software (IBM Incorporate, Armonk, USA). Values are presented as the mean ± standard deviation or median and inter-quartile range. Linear regression was used to model the relationship between dependent and independent variables. We adopted either a two-tailed unpaired t-test (normal distribution) or the Wilcoxon signed-rank test (non-normal distribution). A p-value of less than 0.05 (*p < 0.05) was considered statistically significant.
Results
Validation of the 4D-CBCT System Model
The comparison between PDD measured in water and MC calculations were given in Figure 1. The data were normalized to the dose at 1 cm depth along the central axis of radiation. The deviations between them were less than 1.1%. Figure 2 shows the comparison for cross-line and in-line lateral profiles. For both cross-line and in-line data, the discrepancy between measurements and MC calculations was less than 2.6% in plateau areas, and less than 5% near the fall-off area of the curves.

MC Calculated and Measured Percentage Depth Dose (PDD) Curves of the 4D-CBCT System Using S20 and F1. the top Panel Shows the Normalized PDD Curves, and the Bottom Panel Shows Their Percentage Differences.

MC Calculated and Measured (a) Cross-Line and (b) In-Line Profiles of the 4D-CBCT System Using S20 and F1. The Top Panels Show the Normalized Profile Curves, and the Bottom Panels Shows Their Percentage Differences.
The measured and MC calculated dose in heterogeneous phantoms were displayed in Table 3. Before the kV beam penetrated the tissue mimicking materials (at a depth of 1 cm), the differences between measured and calculated doses were less than 1%. After passing through the tissue-mimicking materials, the deviations were less than 4%.
MC Calculated and Measured Imaging Dose of the CBCT System in Heterogeneous Phantoms.
Imaging Dose from pre-Treatment and in-Treatment 4D-CBCT Scans
Figure 3 shows the typical imaging dose distribution on the CT of a thoracic patient from a single pre-treatment and in-treatment 4D-CBCT scan. They show similar patterns, although the magnitude of the dose differs. Relatively higher doses were found in bone and near the body surface. The statistical distributions of the dose parameters of the typical thoracic OARs from a single pre-treatment or in-treatment 4D-CBCT scan in 8 lung cancer patients are shown in Figure 4. The mean value ± standard deviations of these OAR dose parameters are shown in Table 4. No significant differences were observed in any of these dose parameters between the different breathing modes.

Dose Distribution on a Patient's CT from a Single (a) pre-Treatment and (b) in-Treatment 4D-CBCT Scan, Each Showing the Transverse, Sagittal and Coronal Planes from Left to Right. the red Patch Indicates the Treatment Target.

Dose parameters of the OARs from a Single Pre-treatment or In-Treatment 4D-CBCT Scan.
OAR Dose Parameters from a Pre-Treatment or in-Treatment 4D-CBCT Scan (Mean Value ± Standard Deviation).
The Relationship Between Beam MU or Number of Frames and Imaging Dose from in-Treatment 4D-CBCT Scan
The relationship between beam MU and the number of frames acquired during the in-treatment 4D-CBCT scan for 14 lung cancer patients is shown in Figure 5, along with the linear regression line. There is a relatively significant linear relationship between them. The linear regression equation obtained using least squares method is

Relationship of Number of kV Frames Acquired with in-Treatment Scan as a Function of Treatment Beam MUs.

Relationship of Imaging Dose Parameters of the OARs from In-Treatment Scans as a Function of Beam MUs.

Relationship of Imaging Dose Parameters of the OARs from in-Treatment Scans as a Function of Number of kV Frames Acquired.
OAR Dose Parameters per Frame or Per MU from an in-Treatment 4D-CBCT Scan (Mean Value ± Standard Deviation).
The Increase in OAR Dose Parameters from 4D-CBCT Scans
Figure 8 shows the increase in OAR dose parameters from 5 fractions of pre-treatment and in-treatment 4D-CBCT scans for 8 lung cancer patients when imaging doses were added to the planned doses. For 5, 10 and 25 fractional treatments, the increases in OAR dose parameters are shown in Table 6. For treatments with 5 fractions, the increases in lung V20, lung V5, mean lung dose, mean heart dose, maximum spinal cord dose, maximum esophagus dose and maximum rib dose are 0.04 ± 0.03%, 0.29 ± 0.23%, 76.3 ± 14.4 mGy, 76.5 ± 12.0 mGy, 76.3 ± 20.2 mGy, 70.5 ± 13.1 mGy and 101.2 ± 20.5 mGy, respectively. For treatments with 25 fractions, the increases are 0.21 ± 0.16%, 1.52 ± 1.17%, 380.2 ± 72.1 mGy, 381.6 ± 60.1 mGy, 381.8 ± 100.8 mGy, 352.5 ± 66.5 mGy and 507.4 ± 103.8 mGy, respectively.

The Statistical Distributions of the Increases in OAR Dose Parameters from 5 Fractions of Pre-treatment and In-treatment 4D-CBCT Scans.
The Increases in OAR Dose Parameters When Comparing the Treatment Plan Dose Including and Excluding the Imaging Dose from a Pre-Treatment and an In-Treatment 4D-CBCT Scan Per Fraction.
Figure 9 shows the increase in OAR dose parameters from pre-treatment and in-treatment 4D-CBCT scans per treatment as a function of the number of treatment fractions. All the investigated OARs show a significant linear relationship between the increase in OAR dose parameters and the number of treatment fractions.

The Increase in OAR Dose Parameters from Pre-Treatment and In-Treatment 4D-CBCT Scans Per Treatment as a Function of the Number of Treatment Fractions.
Comparison Between 4D Dose Calculation and AIP CT Dose
Table 7 shows the OAR dose parameters from a pre-treatment 4D-CBCT scan for a patient treated with FB. Both doses calculated using the 4D dose calculation method and on AIP CT are displayed. All the dose parameters calculated using AIP CT differed by less than 0.8% from those calculated using 4D dose method.
OAR Dose Parameters from a Pre-Treatment 4D-CBCT Scan for a Patient Treated with FB.
Discussion
In this study, the imaging dose of pre-treatment and in-treatment 4D-CBCT in patients treated with thoracic radiation therapy and its impact on dose parameters of OARs was evaluated using a model of the XVI imaging system based on MC calculations.
The main reasons for the discrepancies between MC calculated and measured PDDs and profiles could be the slight difference in component dimensions and x-ray tube spectra between the XVI system described in the literature and those used in our institution for the manufacturing error, as well as the measurement errors. The deviations shown in this study are close to similar previous studies.8,14 The discrepancies between the calculated doses in heterogeneous phantoms and the measurements were acceptable.
No significant differences were observed in any of the OAR dose parameters between FB and DIBH patients. This may suggest that the preliminary findings of this study could be applied regardless of the patient's breathing mode.
Preliminary results comparing 4D dose calculations with those calculated using AIP CTs showed that the two methods produce very similar results for OAR dose parameters.
The overall higher magnitude of the dose distribution and OAR dose parameters of pre-treatment 4D-CBCT compared to in-treatment 4D-CBCT can be attributed to the difference in the number of kV frames. The number of kV frames for pre-treatment 4D-CBCT was 1000, while the number for in-treatment 4D-CBCT scans varied between 396 and 869, as shown in Figure 5. The imaging dose can be roughly estimated as proportional to the number of frames. 6 Relatively higher doses were found in bone and near the body surface, both consistent with previous studies for 3D-CBCT.8,12
Yuasa et al reported the average lung and heart mean doses of 143 mGy and 160 mGy, respectively, for lung cancer patients from 4D-CBCT performed with OBI system. 14 Nakamura et al reported median ipsilateral lung mean dose of approximately 30 mGy and heart mean dose of approximately 35 mGy for lung cancer patients from 4D-CBCT performed with Vero4DRT system. 13 The average lung and heart mean doses presented in this study from pre-treatment 4D-CBCT performed with XVI system were 9.5 mGy and 9.7 mGy, respectively, which are much lower than those reported in previous studies. These discrepancies were mainly due to the differences in imaging systems and protocols. For reference, previous studies have shown that the 3D CBCT doses for OBI and XVI systems differ to a large extent. 25 The authors discussed a few reasons for the differences observed. Firstly, the x-ray tubes are located at opposite locations, which determines were the highest doses are deposited on the patient's surface in combination with the acquisition angles used. Secondly, even at similar kVp settings, the x-rays exhibited different levels of filtrations. With such different beams, it was observed that, at similar mAs, more dose is deposited using the OBI unit compared with the XVI unit. The authors didn’t conclude that how different collimator or filtration designs affect the imaging dose. Therefore, this study is only applicable to the XVI system and the scanning protocols used in this paper.
Due to differences in patient geometry, positioning and number of kV frames for in-treatment imaging, the OAR dose parameters showed large inter-patient variations. The variations are in the same range as previous similar studies.13,14 In this study, only 8 patients were selected for MC calculations due to the high cost of computational resources. The clinical findings in this work have limited generalizability due to the small sample size. Future studies could include more patient cases using more computational resources or fast MC algorithms to reduce the variance of the results, and stratified analysis could be used by breathing mode and patient geometry (BMI and anatomical dimensions) to gain further insight. However, variation due to inherent differences between patients cannot be reduced by using more patient samples.
The number of kV frames is considered to be approximately proportional to the imaging dose and has been used for rough imaging dose assessment.4,6 This relationship is also shown in Figure 7. The imaging dose from the in-treatment 4D-CBCT scan is related to the treatment plan and more specifically to the beam MU. For in-treatment 4D-CBCT, the beam MU is negatively correlated with the gantry rotation speed, thus positively correlated with the number of frames acquired (Figure 5), thus positively correlated with the dose parameters to OARs (Figure 6). This relationship can be used in further studies with more patient cases to develop a predictive model for in-treatment 4D-CBCT imaging dose using beam MU as an input. Although the linear relationship between imaging dose and number of frames is more obvious, a predictive model using beam MU allows prediction at the treatment planning stage, which may be more clinically meaningful. At this stage, a rough prediction can be made using the average OAR dose parameters per frame or per MU from an in-treatment 4D CBCT scan shown in Table 5.
Most previous studies have not evaluated the difference between OAR dose parameters with and without imaging dose because it was not considered clinically significant.13,14 While imaging dose is not currently considered in radiotherapy treatment planning, this work provided predictions of the increase in OAR dose parameters in treatment planning when 4D-CBCT imaging dose is added, and demonstrated that the imaging dose may have a clinically relevant impact on OAR dose parameters when used extensively. For 4D-CBCT scans in 5-fraction treatments, the increase in OAR dose parameters were insignificant, as shown in Figure 8. There is a significant linear relationship between the increase in OAR dose parameters and the number of treatment fractions (Figure 9). Therefore, special care should be taken when using 4D-CBCT in conventional radiotherapy, which has more treatment fractions. For example, when using 4D-CBCT with 25 treatment fractions, which is unusual in clinical practice, there will be an apparent average increase of 1.52% in lung V5.
Due to the higher LET of kV radiation, there is an increase in relative biological effectiveness (RBE) of absorbed doses of radiation from kV CBCT sources in reference to megavoltage doses. Previous works have determined the RBE of kV CBCT sources using MC calculations, with results ranging from 1.1 to 1.5. 26 The differences were mainly due to different biological endpoints and beam characteristics, and to the best of our knowledge, no consensus has been reached on the use of the RBE of kV CBCT sources. Therefore, in this study, the RBE of the kV CBCT dose was considered to be the same as the megavoltage dose, following previous studies.3,27
AAPM TG report 180 has provided recommendations for the management of image guidance doses delivered during radiotherapy. 3 It is recommended that imaging dose be considered part of the total dose at the treatment planning stage if the dose from repeated imaging procedures is expected to exceed 5% of the prescribed target dose. In the case of our study, the image dose was well below 5%. However, due to the strong influence of the imaging dose on lung V5 described above, we recommend that the imaging dose should be evaluated when using 4D-CBCT in treatments with 10 or more fractions.
Conclusion
In this study, the imaging dose of pre-treatment and in-treatment 4D-CBCT in patients treated with thoracic radiation therapy and its impact on dose parameters of OARs were evaluated based on MC calculations. The imaging doses of pre-treatment and in-treatment 4D-CBCT showed a similar distribution, but the magnitudes of the doses were different. The investigated OAR dose parameters for in-treatment 4D-CBCT showed a relatively strong linear relationship with the beam MU and the number of kV frames. There is a significant linear relationship between the increase in OAR dose parameters and the number of treatment fractions. For 5-fraction treatments, the imaging dose will likely have minimal clinical impact on OAR dose parameters.
