Abstract
BACKGROUND:
Optimal aerosol delivery methods for spontaneously breathing patients with a tracheostomy remain unclear. Thus, we aimed to assess the impact of nebulizer placement, flow settings, and interfaces on aerosol delivery by using a vibrating mesh nebulizer and a jet nebulizer in line with unheated humidification.
METHODS:
An 8.0-mm tracheostomy tube was connected to the lung model that simulates adult breathing parameters via a collecting filter. Albuterol sulfate (2.5 mg/3 mL) was administered via a vibrating mesh nebulizer and a jet nebulizer, which was placed in line with unheated humidification provided by a large-volume nebulizer, with FIO2 set at 0.28, with gas flows of 2 L/min versus 6 L/min. Nebulizers were placed in line distal and proximal to the lung model by using a tracheostomy collar and a T-piece. Conventional nebulization was tested using a vibrating mesh nebulizer and a jet nebulizer directly connected to the tracheostomy tube bypassing the humidification device. The drug was eluted from the collecting filter and assayed with ultraviolet spectrophotometry (276 nm).
RESULTS:
During in-line nebulizer placement with unheated humidification, the inhaled dose was 2–4 times higher with a gas flow of 2 L/min than 6 L/min, regardless of nebulizer type, placement, or interface (all
CONCLUSIONS:
Aerosol delivery with an in-line vibrating mesh nebulizer and jet nebulizer with unheated humidification was affected by nebulizer placement, interface, and gas flow settings.